| Input ranges | ||||||||||
| ● Voltage ● Current ● Thermocouple ● Resistance thermometer ● Resistance |
Yes
Yes No Yes Yes |
|||||||||
| Input ranges (rated values), voltages | ||||||||||
| ● 0 to +10 V — Input resistance (0 to 10 V) ● 1 V to 5 V — Input resistance (1 V to 5 V) ● 1 V to 10 V ● -1 V to +1 V — Input resistance (-1 V to +1 V) ● -10 V to +10 V — Input resistance (-10 V to +10 V) ● -2.5 V to +2.5 V ● -250 mV to +250 mV ● -5 V to +5 V — Input resistance (-5 V to +5 V) ● -50 mV to +50 mV — Input resistance (-50 mV to +50 mV) ● -500 mV to +500 mV — Input resistance (-500 mV to +500 mV) ● -80 mV to +80 mV |
Yes 100 kΩ Yes 100 kΩ No Yes 100 kΩ Yes 100 kΩ No No Yes 100 kΩ Yes 100 kΩ Yes 100 kΩ No |
|||||||||
| Input ranges (rated values), currents | ||||||||||
| ● 0 to 20 mA — Input resistance (0 to 20 mA) ● -10 mA to +10 mA |
Yes 100 Ω No |
|||||||||
| ● -20 mA to +20 mA — Input resistance (-20 mA to +20 mA) ● -3.2 mA to +3.2 mA ● 4 mA to 20 mA — Input resistance (4 mA to 20 mA) |
Yes 100 Ω No Yes 100 Ω |
|||||||||
| Input ranges (rated values), thermocouples | ||||||||||
| ● Type B ● Type C ● Type E ● Type J ● Type K ● Type L ● Type N ● Type R ● Type S ● Type T ● Type U ● Type TXK/TXK(L) to GOST |
No
No No No No No No No No No No No |
|||||||||
| Input ranges (rated values), resistance thermometer | ||||||||||
| ● Cu 10 ● Ni 100 — Input resistance (Ni 100) ● Ni 1000 — Input resistance (Ni 1000) ● LG-Ni 1000 — Input resistance (LG-Ni 1000) ● Ni 120 ● Ni 200 ● Ni 500 ● Pt 100 — Input resistance (Pt 100) ● Pt 1000 ● Pt 200 ● Pt 500 |
No Yes; Standard/climate 100 MΩ Yes100 MΩ Yes; Standard/climate 100 MΩ NoNo No No |
|||||||||
| Input ranges (rated values), resistors | ||||||||||
| ● 0 to 150 ohms ● 0 to 300 ohms ● 0 to 600 ohms — Input resistance (0 to 600 ohms) ● 0 to 6000 ohms — Input resistance (0 to 6000 ohms) |
No
No Yes 100 MΩ |
|||||||||
| Thermocouple (TC) | ||||||||||
| Temperature compensation | ||||||||||
| — parameterizable — internal temperature compensation — external temperature compensation with compensations socket |
No
No No |
|||||||||
| Characteristic linearization | ||||||||||
| ● parameterizable — for thermocouples — for resistance thermometer |
Yes
No |
|||||||||
| Cable length | ||||||||||
| ● shielded, max. | 200 m; max. 50 m at 50 mV | |||||||||
| Analog value generation for the inputs | ||||||||||
| Integration and conversion time/resolution per channel | ||||||||||
| ● Resolution with overrange (bit including sign), max. ● Integration time, parameterizable ● Basic conversion time (ms) ● Interference voltage suppression for interference frequency f1 in Hz |
13 bit Yes; 60 / 50 ms 66 / 55 ms 50 / 60 Hz |
|||||||||
| Encoder | ||||||||||
| Connection of signal encoders | ||||||||||
| ● for voltage measurement ● for current measurement as 2-wire transducer ● for current measurement as 4-wire transducer ● for resistance measurement with two-wire connection ● for resistance measurement with three-wire connection ● for resistance measurement with four-wire connection |
Yes Yes; with external supply Yes YesYes Yes |
|||||||||
| Errors/accuracies | ||||||||||
| Operational error limit in overall temperature range | ||||||||||
| ● Voltage, relative to input range, (+/-) ● Current, relative to input range, (+/-) ● Resistance, relative to input range, (+/-) ● Resistance thermometer, relative to input range, (+/-) |
0.6 %; ±0.6 % (±5 V, 10 V, 1 to 5 V, 0 to 10 V); ±0.5 % (±50 mV, 500 mV, 1 V) 0.5 %; ±20 mA, 0 to 20 mA, 4 to 20 mA 0.5 %; 0 to 6 kohms, 0 to 600 kohms 1 Kelvin (Pt100, Ni100, climatic; Ni1000, LG-Ni1000, standard; Ni1000, LG- Ni1000, climatic); 1.2 Kelvin (Pt100, Ni100, standard) |
|||||||||
| Basic error limit (operational limit at 25 °C) | ||||||||||
| ● Voltage, relative to input range, (+/-) ● Current, relative to input range, (+/-) ● Resistance, relative to input range, (+/-) ● Resistance thermometer, relative to input range, (+/-) |
0.4 %; 0.4% (±5 V, 10 V, 1 to 5 V, 0 to 10 V); 0.3% (±50 mV, 500 mV, 1 V) 0.3 %; ±20 mA, 0 to 20 mA, 4 to 20 mA 0.3 %; 0 to 6 kohms, 0 to 600 kohms 1 Kelvin (Pt100, Ni100, standard); 0.8 Kelvin (Pt100, Ni100, climatic; Ni1000, LG-Ni1000, standard; Ni1000, LG-Ni1000, climatic) |
|||||||||
| Interrupts/diagnostics/status information | ||||||||||
| Diagnostics function | No | |||||||||
| Alarms | ||||||||||
| ● Diagnostic alarm ● Limit value alarm |
No
No |
|||||||||
| Diagnoses | ||||||||||
| ● Diagnostic information readable | No | |||||||||
| Diagnostics indication LED | ||||||||||
| ● Group error SF (red) | No | |||||||||
| Potential separation | ||||||||||
| Potential separation analog inputs | ||||||||||
| ● between the channels ● between the channels and backplane bus |
No
Yes |
|||||||||
| Isolation | ||||||||||
| Isolation tested with | 500 V DC | |||||||||
| connection method | ||||||||||
| required front connector | 40-pin | |||||||||
| Dimensions | ||||||||||
| Width | 40 mm 125 mm 117 mm |
|||||||||
| Height | ||||||||||
| Depth | ||||||||||
| Weights | ||||||||||
| Weight, approx. | 250 g | |||||||||